Instrumental Measurement:
Most color measurement instruments have not been designed to measure the
effects of fluorescence. Color instruments found in most textile companies are
spectrophotometers that are either 450 or sphere diffuse 8 geometry
instruments. They are what we refer to as single monchromator instruments where
light reflected or transmitted from the sample is broken into its
respective wavelengths and presented to the analyzer. The analyzer calculates
percentages of this light at its designated bandwidths and displays the color
curve. The light from the sample is referred to as polychromatic (white light)
and the instrument as a polychromatic illuminating spectrophotometer. This is
an efficient, accurate, and repeatable method of measuring non-fluorescent
samples.
As we know, fluorescent samples absorb light in other regions of the spectrum
and then re-emit in the visible area of the spectrum. Ideal instruments for
measuring this effect would be those that have the ability to separate the base
reflectance from the amount of reflectance caused by the fluorescence. Such
instruments do exist but they are expensive and not common to the industry.
They are dual (2) monchromator a system where the first monchromator breaks the
illuminant into its individual wavelengths and the second monchromator
receives the light from the sample (at its specific wavelength) and presents it
to the analyzer. Output data can separate the base reflectance from the
fluorescent reflectance and allow better analysis of the effects of optical
brightners and fluorescent dyes and/or pigments. Again, these instruments are
expensive, slow in measurement time, and not common to the other applications
within the textile industry.
When using conventional spectrophotometers for fluorescent measurements and
comparing their results to other instruments, the following conditions should
be considered:
- Geometry of Instrument.
- Measurement Area
- Light Source of Instrument
- Illumination Filtered to D65
- Calibration of the UV Component
Instrument Geometry: Measurement Area:
Two types of geometry found in conventional spectrophotometers are sphere (8)
and 45/0 or 0/45. Both geometries can be used to measure fluorescent samples
and if comparing data from two (2) instruments it is important to insure that
they have identical geometry. If not, measurement data may be different and may
not be compatible. This is true not only with fluorescent samples but also with
any measured samples.
Another variable required for consistency is that of aperture or area of
measurement on the instrument. As an example, an 8mm circular aperture will provide reflectance energy from an 8mm area of the sample. When measuring fluorescent
standards/samples the data collected is a result of the measuring area. If the
measuring area is changed to 25.4mm the data collected may not correlate to the
data collected from the smaller measuring area. Reflectance from the
fluorescent sample may be higher or lower in UV content relative to the area of
measurement.
Light Source: Illuminant D65:
Two important specifications are necessary relative to the light source of the
spectrophotometer. One, in order to meet CIE specifications for the measurement
of fluorescent samples it is necessary for the illuminant to contain ample
levels of UV energy. Secondly, it is necessary for the light source illuminating
the sample to be filtered to D65. Some instruments use quartz or tungsten light
sources that lack ample UV energy to excite the whitening or fluorescent agents
within the sample. Whatever UV energy they do contain begins to diminish early
into the life cycle of the lamp. Because of the low amount of UV energy emitted
by these type lamps they are more difficult to filter to D65.
Instruments that have pulsed-xenon light sources contain a high amount of UV
energy and they are easier to filter to D65. Therefore, this type source
provides the UV power to simulate D65 daylight.